Beam | |
---|---|
Energy | 18-45 MeV |
Undulator | |
period (λu) | 60 mm |
mo. periods | 46 |
KRMS (range) | 0.4–1.7 |
Rayleigh range Mol. beam | 55 mm |
Rayleigh range FTICR | 82 mm |
Type | Quasiperiodic PPM |
FELICE is the beam line of FELIX operating from 5–100 µm and allowing intra-cavity experiments.
Procedures
Shutdown Procedure
- File:Shutdown FELICE quick guide.docx: how to shut down FELICE
- File:Shutdown FELICE.docx: how to shut down FELICE
Attainable wavelengths as function of beam energy
Gap → ↓ Energy (MeV) |
23 | 25 | 27,5 | 30 | 32,5 | 35 | 37,5 | 40 | 42,5 | 45 | 47,5 | 50 |
---|---|---|---|---|---|---|---|---|---|---|---|---|
15 | 143,22 | 122,03 | 101,29 | 85,50 | 73,50 | 64,36 | 57,40 | 52,11 | 48,08 | 45,01 | 42,67 | 40,89 |
17,5 | 105,22 | 89,65 | 74,41 | 62,82 | 54,00 | 47,28 | 42,17 | 38,28 | 35,32 | 33,07 | 31,35 | 30,05 |
20 | 80,56 | 68,64 | 56,97 | 48,10 | 41,34 | 36,20 | 32,29 | 29,31 | 27,04 | 25,32 | 24,00 | 23,00 |
22,5 | 63,65 | 54,23 | 45,02 | 38,00 | 32,66 | 28,60 | 25,51 | 23,16 | 21,37 | 20,00 | 18,97 | 18,18 |
25 | 51,56 | 43,93 | 36,46 | 30,78 | 26,46 | 23,17 | 20,66 | 18,76 | 17,31 | 16,20 | 15,36 | 14,72 |
27,5 | 42,61 | 36,31 | 30,13 | 25,44 | 21,87 | 19,15 | 17,08 | 15,50 | 14,30 | 13,39 | 12,70 | 12,17 |
30 | 35,81 | 30,51 | 25,32 | 21,38 | 18,37 | 16,09 | 14,35 | 13,03 | 12,02 | 11,25 | 10,67 | 10,22 |
32,5 | 30,51 | 25,99 | 21,58 | 18,21 | 15,66 | 13,71 | 12,23 | 11,10 | 10,24 | 9,59 | 9,09 | 8,71 |
35 | 26,31 | 22,41 | 18,60 | 15,70 | 13,50 | 11,82 | 10,54 | 9,57 | 8,83 | 8,27 | 7,84 | 7,51 |
37,5 | 22,92 | 19,52 | 16,21 | 13,68 | 11,76 | 10,30 | 9,18 | 8,34 | 7,69 | 7,20 | 6,83 | 6,54 |
40 | 20,14 | 17,16 | 14,24 | 12,02 | 10,34 | 9,05 | 8,07 | 7,33 | 6,76 | 6,33 | 6,00 | 5,75 |
42,5 | 17,84 | 15,20 | 12,62 | 10,65 | 9,16 | 8,02 | 7,15 | 6,49 | 5,99 | 5,61 | 5,32 | 5,09 |
45 | 15,91 | 13,56 | 11,25 | 9,50 | 8,17 | 7,15 | 6,38 | 5,79 | 5,34 | 5,00 | 4,74 | 4,54 |
47,5 | 14,28 | 12,17 | 10,10 | 8,53 | 7,33 | 6,42 | 5,72 | 5,20 | 4,79 | 4,49 | 4,26 | 4,08 |
50 | 12,89 | 10,98 | 9,12 | 7,70 | 6,61 | 5,79 | 5,17 | 4,69 | 4,33 | 4,05 | 3,84 | 3,68 |
Cavity Length
Pay attention before changing cavity length When you feel you have to change the cavity length here, please be sure that the US and DS mirrors are properly homed, i.e. that the value you commit is really valid. Make sure that the cavity length is preset to zero at the home position of both mirrors! |
switches | Molecular Beam position (µm) DS end mirror |
FTICR position (µm) US end mirror |
remarks and dates last checked |
---|---|---|---|
home position | 0 | 0 | by definition |
cavity zero | 271 | -1217 | MB 08/04/2021, FTICR 19/05/2021 |
negative limit | -1262 | -2246 | MB 03/09/2020, FTICR 09/09/2020 |
positive limit | 858 | 1130 | MB: 858 is the maximum reachable, after that the motor probably loses contact with the stage (03/09/2020), FTICR: 1330 (09/09/2020) |
(i.e. the stated cavity zero position is relative to the home positions).
Debug info
On the upstream side, the homeswitch is active towards the negative limit switch, and inactive towards the positive limit switch. A positive movement of the stepper motor motor corresponds to a longer cavity length. Because of the nature of piezomotors, the step size is not equal in both directions. At the position of the home switch +10000 steps corresponds to +6 µm, and -10000 steps is ca. -23 µm.
Mirror Settings
After changing from Molecular Beam to FTICR and vice versa:
- Cavity: home both US and DS
- CL zero: set to known last value
- X/Y-positions for all mirrors need to be homed
mirror | Molecular Beam position (steps) | FTICR position (steps) |
---|---|---|
DS | H -78 V -125 |
H 78 V -330 |
US | H -133 V -129 |
H 60 V 0 |
Angle | H 182 V -76 |
H -186 V 239 |
End | H -36 V 56 |
H 232 V -17 |
Last updated | 2025-02-24 | 2025-02-18 |
Diagnostic station
Software for diagnostics
Spectrum analyzer
For the Felice spectrum analyzer there are two GUI interfaces "FELICE spectroclient.exe" and "sa_flc_pyro_gui_v101.exe". The program "sa_flc_pyro_gui_v101.exe" is only needed in special cases.
- For correct working of showing the wavelength use the even+odd setting from the pyro_gui interface. This will ensure the correct dispersion.
- startCh needs an offset of about 5 channels due to strange outlines
- endCh' chould be set normally at 256, if selected in the pyro GUI to even or odd the channels should be set to 128 but at the moment not advisable.
- FSSP (Felix super stabilization program): if switched on will subtract a baseline.
- CalibrCycleON: switched on will try to make an offset array of each channel to zero it. When done the switch will turn gray again.
- CalibrOffsetOn: if set ON this will use the offset array to compensate the baseline to zero.
Server program (by Guus Tielemans) runs on Feloce.
Required connections:
- RS232 connection to control Princeton Instruments Grating spectrometer (male DB9 feedthrough on DS)
- Rijnhuizen built array detector (female DB9 feedthrough on DS)
\\isolator\flare-rinst\Felix\Felix_prg\SA_felice\FELICE spectroclient.exe
Molectron power meter
Server program (by Guus Tielemans) runs on Feloce.
Required connections:
- TTL trigger for power meter
- GPIB-ethernet controller connected to Molectron controller
- Power head connected to input Molectron (LEMO feedthough connection on DS)
\\felixdisk\msd-lab-clusters\labview-programs\PowerscanFELICE\v200220\PowerscanFELICE.exe
Note: this is compiled in Labview 2017, due to NI support for the ethernet Gpib controller being discontinued in later versions.
Representative output energies
Molecular beam
Power measurements done with Coherent EPM1000 power meter, in combination with Molectron 750LF-1 power head. Responsitivy 10.8 V/J
FTICR
Notes on how to convert the outcoupled pulse energy into the macropulse energy within the cavity
Timing & Diagnostic connections on the patch panel in the FELICE hall
- Felice cavity dumping requires a phase-stable trigger pulse: route E5 through the black RF locking box (named _Henny_ in the diagram above).
- The FTICR won't run with short trigger pulses, requires at least 10 us wide pulse
- to enable timer E5 to have long pulse lengths, it should be configured as follows
Troubleshooting
- Is the Pyro hooked up well?
- Downstream or Upstream mirror can't be changed (select mirror => Go remains greyed out)? Soft restart of the Subsystem will do the trick. (This button is greyed out for the first few minutes after start of the GUI.)
- If FELICE does not trigger even though all lights are green, check to see if there is either a) a terminator or b) a BNC cable connected to the top-left BNC connector (M110) on the FELICE patch panel behind the cluster source. It should be open.
- FTICR specific:
- Is the Faraday cup out? (You should be able to see the alignment HeNe on screen 7S80(near end mirror).)
- FELICE Molecular beam specific:
- Is the cavity-dump slab out? (You should be able to see the alignment HeNe on screen 7S60 (near end mirror).)
- Is the experiment not too high? Check for straightness of the bellows between experiment and mirror vessels. Check bestec PC--> Motion control -->V. Value vertical should be between -32 and -28 mm.
Pictures
References and notes
Documentation
- Overall
\\felixdisk.science.ru.nl\ffg2\Technical Documentation\FELICE
- Tech specs
\\felixdisk.science.ru.nl\ffg2\Technical Documentation\FELICE\Technical Specifications.doc
- Reconstruction of intracavity power from measured outcoupled pulse measurement
\\felixdisk.science.ru.nl\ffg2\Technical Documentation\FELICE\Intracavitypower.xls